We are using cookies on this website. We assume your consent, as you are making use of this website. A cookie is a simple small file sent with the pages of this website and stored by your browser on the hard disk of your computer. The stored information can be resent to our servers the next time you visit the website. More information
Hide this message
Sysmex Nederlands
Sysmex Europe
Service
Support
Other Sites
Menu

Percentage of hypo-haemoglobinised red cells (HYPO-He) and hyper-haemoglobinised red cells (HYPER-He)

What are HYPO-He and HYPER-He used for?

Together with other parameters, HYPO-He is used in RBC indices to distinguish between iron defiency anaemia and thalassaemia.

The use of HYPO-He as well as MicroR is recommended in nephrology guidelines such as the European Best Practice Guidelines (EBPG), National Kidney Foundation Kidney Disease Outcome Quality Initiative (NKF KDOQI).

How are HYPO-He and HYPER-He determined?

HYPO-He and HYPER-He are parameters analysed in the reticulocyte (RET) channel. They are derived from the haemoglobin content of all mature RBC (RBC-He), which can be calculated based on the high-angle forward scatter (FSC). The FSC signal is converted into picograms (pg) using a proprietary algorithm and in healthy individuals the resulting RBC-He value is comparable to the mean corpuscular haemoglobin (MCH) value. Subsequently, this RBC-He value is used to determine HYPO-He and HYPER-He.

HYPO-He is the percentage of RBC with cellular haemoglobin content lower than 17 pg, whereas HYPER-He is the percentage of RBC with cellular haemoglobin content higher than 49 pg.


Figure: Each cell is plotted in the RET scattergram based on its fluorescence intensity (x-axis) and its high-angle forward-scattered light signal (y-axis), which reflects characteristics of both cell size and cellular content. The left panel shows a sample from a healthy individual with HYPO-He less than 1% and the right panel shows a sample with HYPO-He more than 60%

HYPO-He and HYPER-He are diagnostic parameters on XN-Class analysers that have the RET channel.*

*In the XT-4000i and XE-5000 analysers these are research parameters that are derived from the RET channel and named %HYPO-He and %HYPER-He. Research parameters should not be used for in vitro diagnostics.