We are using cookies on this website. We assume your consent, as you are making use of this website. A cookie is a simple small file sent with the pages of this website and stored by your browser on the hard disk of your computer. The stored information can be resent to our servers the next time you visit the website. More information
Hide this message
Sysmex Nederlands
Sysmex Europe
Service
Support
Other Sites
Menu
Sysmex Nederlands Academy Knowledge Centre Sysmex Parameters Fragmented red blood cells (FRC*)

Fragmented red blood cells (FRC*)

Sysmex analysers use the fluorescence flow cytometry method in the reticulocyte channel to measure fragmented red blood cells (FRC% and FRC#) as research parameters.

A specific area below the RBC area in the RET scattergram is used for identification of fragmented red blood cells. Due to the absence of nucleic acids in red blood cells the intensity of the measured side fluorescence signals (SFL) is extremely low. In addition, the high-angle forward scatter (FSC) is lower than that of intact red blood cells.

Fragmented red blood cells (FRC*)
Figure: Each cell is plotted in the RET scattergram based on its fluorescence intensity (SFL on x-axis) and its high-angle forward scatter (FSC on y-axis), which reflects characteristics of both cell size and cellular content. The triangle indicates the detection area for fragmented red blood cells (FRC)

Fragmented red blood cells are generally the consequence of mechanical damage, usually in the context of turbulent blood flow or contact with a pathologically altered endothelium, the latter occurring most commonly in the microvasculature. These abnormal shear forces destroy the integrity of the red blood cells, producing cell remnants that appear as ‘helmets’ (cells with two tapered and hornlike projections on either end) and other odd shapes when viewed under a microscope.

Figure: Peripheral blood smear with ‘helmet’ cells (arrows)

* FRC% and FRC# are research parameters obtained in the RET channel on XN-Series and X-Class analysers (XT-4000i and XE-5000). Research parameters should not be used for in vitro diagnostics.