Size Range
0.3 to 25μm
Size Channels
Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm
Custom binning available
Flow rate
0.1 CFM (2.83 LPM)
Concentration Limit
4,000,000 Particles/ft³ @ 5% coincidence loss
Battery Run Time
PC-4000 10+ hours, PC-4005 8+ hours (typical)
Light Source
Long life laser diode
Counting Efficiency
50% @ 0.3 μm; 100% for particles >0.45 μm per JIS
Zero Count <1 count/5 minutes (<2 particles/ft³) (per ISO 21501-4 & JIS)
Count Modes
Automatic, manual, beep, cumulative/differential counts or mass concentration
Count Alarms
1 to 9,999,999 counts
Carbon Dioxide (Optional)
0 to 5,000 ppm ±65ppm ±3% rdg. (PC-4005 only)
Relative Humidity/Temperature (Optional)
0 to 98 %RH ±2%RH, 32 to 122 °F (0 to 50 °C) ±1ºF (0.5ºC)
TVOCs (Optional)
0 to 50ppm range, <5ppb L.O.D. (PC-4005 only)
Display
4.3” (10.9 cm) WQVGA (480x272) color touch screen
Vacuum Source
Internal pump with automatic flow control
Filtered Exhaust
Internal HEPA filter
Number of Channels
6 for particulate, plus 2 for optional TRH (PC-4000), plus 2 additional for optional sensors on the PC-4005
Custom Size Channels
Calibration for custom particle size channels available
Audible Alarm
Adjustable built-in alarm
Battery
Removable Li-ion
Battery Recharge Time
4 hours
Recipes
50 user-configurable recipes
Communication Modes
Serial (for interface with AdvancedSense Pro, WP, Tablets and data download to WolfSensePC), USB for calibration and firmware updates
Optional Communication Modes
RS-232/RS-485 Modbus and Ethernet hardware is standard. Wireless 802.11 b/g (with antenna) is optional. Interface with these comm modes requires optional software
Alarm
Alarms on counts for all particle sizes, low battery, sensor failure, environmental sensors and flow
Standards
ISO 21501-4 and JIS B9921
Factory Calibration
Recommended annually. NIST traceable (cert optional)
External Surface
Stainless steel
Dimensions (L x W x H)
13.3 cm x 10.5 cm x 21 cm
incl. barb fittings (9.5”, 24 cm H w/isokinetic inlet)
Weight
1.8 kg
Included Accessories
Quick start guide, operating manual on USB flash drive, isokinetic probe, purge filter, WolfSense® PC data download and report generation software, USB cable, power supply (spec US, UK or Euro plug)
Additional Optional Accessories
Calibration hood, printed manual, spare battery, spare external battery charger and isokinetic probes
Buffer Memory
40,000 sample records (rotating buffer) including particle count data, environmental data, locations and times.
Scrollable on screen or printout
Sample Locations
Up to 1,000 locations 20 characters long (unlimited locations 256 characters long, when interfaced to AdvancedSensePro/WolfPack/DirectSense)
Sample Time
1 second to 99 hours
Power
110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions
41º to 104ºF (5º to 40ºC)/20% to 90%RH non-condensing
Storage Conditions
32º to 122ºF (0º to 50ºC) / Up to 98%RH non-condensing
Warranty
2 years limited (1 year on internal pump and optional TVOC, CO2 sensors)