Sysmex Nederlands
Sysmex Europe
Service
Other Sites
Menu

Thrombocytopenia and detection of cytoplasm fragments from monoblasts

Thrombocytopenia and detection of cytoplasm fragments from monoblasts

Thrombocytopenia and detection of cytoplasm fragments from monoblasts in AML-M5A. Due to their similar size the cytoplasm fragments can produce a falsely elevated platelet count in impedance measurement on certain haematology analysers. Here the initial count was 32,000/μL. The correct platelet value was obtained from the optical channel; it was 7,000/μL. Cytoplasm fragments (C) and a platelet (P) are marked.

Enjoy our Media Centre
All
  • All
  • Documents
  • Podcast
  • Images
  • Vidéos
Growing your knowledge
Visit our Academy
Growing your knowledge Visit our Academy
See our library of smears https://www.sysmex.nl/nl-fr/library/image-gallery.html
Growing your knowledge Visit our Academy
See our library of smears https://www.sysmex.nl/nl-fr/library/image-gallery.html
Growing your knowledge Visit our Academy