Malvern Kinexus pro+

  • All modes of rheological operation – stress control, shear rate control and direct strain controlled oscillation at demand strain amplitude for accurate control of sample strain history
  • Exceptional vertical travel and gapping capabilities with ultra-responsive and highly sensitive normal force system for class-leading performance
  • Unique rSpace software interface that offers total flexibility of test set-up - from sequence-driven Standard Operating Procedure (SOP)-type functionality to fully customizable test design for new and advanced research capabilities

Documents

All
  • All
  • Brochure
Documents

Research grade rheometer for complex fluids characterization

Targeted at the rheological characterization of dispersions and other complex fluids and soft solids, the Kinexus pro+ rheometer has unprecedented dual-action capabilities for both shear and vertical testing. A modular rheometer with true ‘plug and play’ functionality for all measuring systems and environmental control units, Kinexus also enables pioneering Standard Operating Procedure (SOP) driven rheological testing.

Request a quote

Video

Video

Specifications

Specifications

Modes of operation
Direct strain control, Shear rate control, Shear stress control

Torque range
5.0nNm to 225mNm (Viscometry - Controlled Rate and Controlled Stress)  1.0nNm to 225mNm (Oscillation - Controlled Strain and Controlled Stress)

Torque resolution
0.1nNm

Position resolution
<10nrad

Angular velocity range
1nrads-1 to 500rads-1

Step change in strain
<10ms

Frequency range
6.28µrads-1 to 942rads-1 (1µHz to 150Hz)

Motor inertia
12µN.m.s2.

Normal Force range
0.001N to 50N

Normal Force resolution
0.5mN

Normal Force response time
<10ms

Vertical lift speed
0.1µms-1 to 35mms-1

Vertical lift range (measurable)
230mm

Gap resolution (over full vertical lift range)
0.1µm

Temperature range
-40°C to 200°C (Peltier plate and active hood cartridges)
-30°C to 200°C (Peltier cylinder cartridge)
0°C to 300°C (High temperature cartridge)

Temperature resolution
0.01°C

Dim
ensions
490mm (W) x 485 (D) x 680 (H)

Weight
47kg