SPECTRO ARCOS ICP-OES

The SPECTRO ARCOS ICP-OES excels in industrial and academic applications for the most advanced elemental analysis of metals, chemicals, petrochemicals, and other materials.

MultiView
The periscope-free MultiView mechanism lets an operator literally "turn" a radial-view instrument into an axial-view device, or vice-versa, in 90 seconds or less. Visit the MultiView Q&A section in the ARCOS Resource Center for answers to frequently asked questions.

Design
The design of the SPECTRO ARCOS ensures exceptionally low operating costs over a long, reliable service life. And it packs a modern, ergonomic chassis with proven features such as no-purge UV-PLUS sealed gas purification technology, no-external-cooling OPI-Air Interface.

Specifications
Polychromator Thermally stabilized to +15°C ±0.5°C
ORCA hollow section design in triple Paschen-Runge mounting
750 mm focal length
Holographic concave master grating: 2 x 3600, 1 x 1800 g/mm
Grating material Zerodur
MgF2 optical components
Full 1st order wavelength coverage
Wavelength range: 130 (160) - 770 nm
Detector 32(29) linear CCD arrays
3648 pixels per array
Optical/pixel resolution: 130(160)-340 nm 8.5 pm/3 pm, >340 nm 16 pm/6 pm
Thermally stabilized optical system (+15°C ±0.5°C)
Dynamic range up to 9 orders of magnitude
Shortest integration time: 0.1 ms
Shortest measurement time for one analysis: 2 s
UV System UV-PLUS system
Gas (argon) filled hollow section, sealed, no consumable purge gas required
Dual window entrance optics, easy to maintain
Automatic gas purifying system
Lifetime of purifying cartridge: 24 months
RF Generator LDMOS solid-state design
Frequency: 27.12 MHz, free running type
RF power output: 0.5 to 2.0 kW
Power efficiency: >85%
Power stability: ≤0.1%
Air cooled (no external cooling required
Dimensions 1068 (H) x 1582 (W) x 756 (D) mm
Weight Approx. 240 kg
Environmental Conditions Room temperature: 15-35°C
Exhaust System Requirements Capacity: 2 x 300 m3 per h (177 cft/min), separately adjustable between zero and maximum
Electrical Requirements 230 VAC ± 5%, 50/60 Hz
Approx. 4.0 kVA power consumption
30-32 A instrument required line protection (slow blow fuse)
Contact
You have a question and would like to get in touch with our experts? We look forward to receiving your enquiry.